Environmental Growth Room
Feature
- Polyurethane Insulated panel structure with stainless steel interior and powdercoated steel exterior
- Assembled and disassembled construction structure for space saving shipment
- Stainless steel plate 3.0 stickness for floor
- All-in-One Controller that integrates Monitoring and Recording Functions
- Built-in refrigeration unit of chamber
- Exterior connection refrigeration unit at out door
- Air cooled condensing system
- Water cooled condensing system
- Reliable Hardware and Software
- 5.7 inch TFT LCD touch screen with dust – proof controller
Steel Fabricated Structure
- Real time monitoring and data acquisition
- Display graph of saved data
- Graph and setup patterns
- 120 Patterns / 1200Segments
- Monitoring on PC Software
- Remote operation & editing programs
- Graph & Data acquisition
- Converting saved data to spreadsheet
Specification
MODEL |
LWG-3000RT |
LWG-4000RT |
LWG-6000RT |
Dimensions Inner (wxhxd mm) Outer |
1200x1200x2000 2000x1600x2540 |
1200x1700x2000 2000x2130x2540 |
1800x1700x2000 2745x2133x2540 |
Capacity |
2880â„“ |
4080â„“ |
6120â„“ |
Internal usable area |
1.44
m² |
2.04
m² |
3.06
m² |
Temperature |
Range
4°C to 60°C (Light Off) / 10°C to 60°C (Light On) Set
Accuracy ±0.01°C / Uniformity : ±1.0°C at 25°C |
||
Humidity |
Range
30% to 95% RH (Light off) / 30% to 85% RH (Light on) Set
accuracy ±1% RH / Uniformity : ±2% to 3% at 24% Electronic
Sensor (free of drying a wet and good response time) Evaporation
by Fan with Heater and automatic water supply |
||
Controller |
Microprocessor
PID Multi-Function Controller for recorder data, program and alarm
information |
||
Display |
Graphic
Display of set point and actual performance |
||
Illumination |
0 to
35000 Lux (Option) |
||
Work Light |
Vapor
proof fluorescent lamp |
||
Material Interior Exterior |
Stainless
steel with polished finish Epoxy
powder coated EGI Steel |
||
Door |
Two
doors with four of viewing windows |
||
Refrigeration System |
Selectable
Air cooled of water cooled condenser / Built-in at chamber / of External
Installation |
||
Air Flow |
2
sets of sirocco fan (200W), HEPA filtered air supply |
||
Safety |
Switch-off
after alarm for overheating, water empty, current leak, over current breaker |
||
Electric Supply |
110V
60Hz or 220V 50/60Hz or 3800V 50/60Hz, 3Phase / 4Wires |