Giant Environmental Chamber LGC-3301RT

 
LGC-3301RT

Giant Environmental Chamber

Feature

Heavy – Duty Mechanical Design
  • Manufactured for completed sealed chamber structure
  • Easy to move and install with sturdy foot master
  • HEPA filtered clean air is supplied into the chamber for culture and growth, installed at roof top
  • Perfect uniformity air circulation chamber structure, with air velocity control technology

Non-CFC/HCFC Refrigeration System
CFC / HCFC – free Hermetically sealed compressor and eco – friendly refrigerant gas are used for compliance with the New Environmental Regulation

Selection of refrigeration units for Installation
  • Built-in refrigeration unit of chamber
  • Exterior connection refrigeration unit at out door
  • Air cooled condensing system
  • Water  cooled condensing system

All-in-one Controller that integrates Monitoring and Recording Functions
  • Reliable Hardware and Software
  • 5.7 inch TFT LCD touch screen with dust – proof controller

Automatic shelf moving mechanism
  • Motorized lift up and down shelf in accordance with the height of products, plants and samples

Steel Fabricated Structure

Microprocessor Programmer PID Control with PC Interface (Option)
  • Real time monitoring and data acquisition      
  • Display graph of saved data       
  • Graph and setup patterns
Controller
  • 120 Patterns / 1200Segments
  • Monitoring on PC Software
  • Remote operation & editing programs
  • Graph & Data acquisition
  • Converting saved data tospreadsheet

Specification

MODEL

LGC-3301RT

LGC-3302RT

LGC-3303RT

Dimensions  Inner

(wxdxh mm) Outer

1850x810x1550

2870x1150x2050

2000x1000x1550

3420x1300x2350

2500x1400x1550

3620x1700x2050

Capacity

2332â„“

3000â„“

5425â„“

Internal usable area

1.5 m²

2 m²

3.5 m²

Temperature

Range 4°C to 60°C (Light Off) / 10°C to 60°C (Light On)

Set Accuracy ±0.01°C / Uniformity : ±1.0°C at 25°C

Heater

4.5kW

4.5kW

4.5kW

Compressor

3 / 2HP

3 / 2HP

2HP

Humidity

Range 30% to 95% RH (Light off) / 30% to 85% RH (Light on)

Set accuracy ±1% RH / Uniformity : ±2% to 3% at 24%

Electronic Sensor (free of drying a wet and good response time)

Evaporation by Fan with Heater and automatic water supply

Display

Graphic Display of set point and actual performance

Controller

Microprocessor PID Multi-Function

Illumination Range

               Lamp

               Bank

               Control

0 to 25000 Lux (Option)

Each tiers FL 20W x 8EA

Ceiling of each tier

0 to 35000 Lux (Option)

Each tiers FL 40W x 8EA

Ceiling of each tier

0 to 35000 Lux (Option)

Each tiers FL 40W x 8EA

Ceiling of each tier

Timer for each tiers and programming

Material      Interior

  Exterior

Stainless steel with polished finish

Epoxy powder coated EGI Steel

Door

Two doors with four of viewing windows

Refrigeration   System

Selectable Air cooled of water cooled condenser / Built-in at chamber / of External Installation

Air Flow

2 sets of sirocco fan (200W), HEPA filtered air supply

Safety

Switch-off after alarm for overheating, water empty, current leak, over current breaker

Electric Supply

110V 60Hz or 220V 50/60Hz or 3800V 50/60Hz, 3Phase / 4Wires

 


#Labtech, #Daihanlabtech, #Trias, #pt.trias, #alatlab, #laboratorium, #alatkesehatan, #Waterbath, #Daihan, #alatindustri, #Fume Hood, #Giant Environmental Chamber,

Related product you might see:

Share this product :

 
Support : Creating Website | PT.TRIAS NATHOMI CHEMINDO | www.triasnathomichemindo.com
Copyright © 2011. PT.TRIAS NATHOMI CHEMINDO - All Rights Reserved
Template Created by Creating Website Published by www.triasnathomi.com
Proudly powered by TRANALAB