Feature
Sterilization
temperature 110°C to 135°C EN-13060-2 class B sterilization cycle
Pre Vacuum-water filling-heating-sterilizing-ventilation-post
vacuum drying-end cycle
Pre-vacuum
and post vacuum Drying system
Pre-vacuum for effective air removal
that results in increasing sterilization performance
Post-vacuum for drying with remaining
steam removal
Microprocessor
PID program controller with self auto-tuned and LCD Monitor
7 pre-programmed cycle as 2 wrapped, 2
unwrapped liquid, Bowie-dick and 1 user programmable course
Full
automatic operation system
Easy operation with full automatic process
cycle by microprocessor controller from
beginning to end cycle
Self-diagnostic
and Safety System
In case of any malfunction, it alarms
and displays error codes for easy maintenance, being stop working with
automatic ventilation process for safety purpose
Flushing
Cycle Sterilization system
Can be supplied as optional order
Air
filtration cooling system
For the process of drying cycle, 0.2µm
HEPA filtered clean air flows into the chamber in order to protect air
contamination and cool down below 60°C
Safe
and Efficient water supply system
Automatic water supply mechanism which
supplies a fixed volume of water to the chamber for maximum efficiency
Automatic
and manual door lock system
When the sliding handle is entered into
the locking ring, the door automatically locks after two second. The door can
be manually opened and locked using the locking wheel system
Space
saving compact design
Maximize working capacity with square
chamber and minimize foot space with dedicated engineering design.
Safety
Mechanism
1. Over pressure release valve
2. Over temperature protector with
Thermostat
3. Higher temp set protector by Digital
high alarm limit
4. Visible and Audible alarm system
5. Automatic dual door locking system
Programmed Sterilization Cycle
MODEL
|
Temperature (°C)
|
Pressure (kgf / cm2)
|
Sterilizing Time (min)
|
Drying Time (min)
|
Unwrapped 1
|
121
|
1.1
|
20
|
30
|
Unwrapped 2
|
134
|
2.1
|
5
|
30
|
Wrapped 1
|
121
|
1.1
|
30
|
30
|
Wrapped 2
|
134
|
2.1
|
15
|
30
|
Liquid
|
121
|
1.1
|
20
|
0
|
Bowie Dick
|
134
|
2.1
|
4
|
5-10
|
User Program
|
100-135
|
0.1-2.3
|
4-30
|
0-60
|
MODEL
|
LAC-1011SV
|
LAC-1021SV
|
LAC-1031SV
|
||
Dimensions
Inner
(W x D x H
mm) Outer
|
410 x 560 x 460
200 x 300 x 200
|
490 x 655 x 460
270 x 405 x 230
|
510 x 690 x 545
300 x 455 x 300
|
||
Capacity
|
12â„“
|
25â„“
|
40â„“
|
||
Heater
|
2kW
|
2.5kW
|
2.5Kw
|
||
Chamber Type
|
Reinforced and electro polished square chamber
|
||||
Material
|
Stainless Steel 304 standard (STS 316 L option)
|
||||
Door lock mechanism
|
Automatic wheel inter locking with one touch sliding handle
|
||||
Vacuum Pump
|
Built-in diaphragm vacuum pump 25 LPM capacity
|
||||
Controller
|
Microprocessor Program Controller with LCD Monitor
|
||||
Sterilization Temp Range
Pressure Range
Course
|
110°C ~ 135°C
0.4 ~ 2.3kgf/cm2
Unwrapped 1&2 / Wrapped 1&2 / Liquid / Bowie-Dick /
User program mode
|
||||
Reservior
|
6â„“ stainless steel tank
with top wazer filling inlet and silicon closer
|
||||
Safety
|
Over pressure release valve, Over temperature protector,
Visible and Audible alarm system, Automatic dual door locking
system
|
||||
Tray
|
180 x 260 mm
|
250 x 360 mm
|
290 x 400 mm
|
||
Perforated tray (Standard) / Wire tray (Option)
|
|||||
Electric Supply
|
110V 60Hz or 220V 50/60Hz
|