Bio Hazard Safety Cabinet Bench Top Class Ⅱ Type B2
In compliance with ISO 14644-1, ISO Class 3 (Class 10) standards
- ULPA Filters
That provide 99.999% efficiency for particle sizes of 0.3um
are installed for both supply and exhaust filters
- Pressure Gauge
Is installed as standard to check filter lifetime status
- Temperature Safety Glass Door with Safety Alarm
In case the door is opened above the proper height
- High Efficiency Dual Inlet Blower
Is installed to provide reduced noise and vibration
- Utility Device
Includes npt ¼ air valve, gas valve, 110V or 220V electric
socket (selectable), DPO test valve, UV lamps and FL lamps are installed as
standard
- Stainless Steel Single Place Dished Work Space
For easy cleaning and to prevent contaminants
- Arm Rest
Provides comfortable user working posture
- Paper Catch
Is installed as standard
- Ergonomic Design
7” angled ergonomic structure allows easy reach and viewing
while also minimizes stress on user’s neck and arm
Microprocessor Digital Controller with LCD Display
- LCD Display
- Filter Lifetime Check
- Alarm, Lamp on/off, Fan on/off status indicator
- The controller enables
- Velocity setting, alarms filter lifetime check, UV&FL lamp on/off, UV lamp timer, room temperature, present date and time.
Air Flow
100% of the air is exhausted
- External blower should be separately installed for 100% exhaust.
Optional Accessories
- Base Stand with Castors
- Frame Type Stand with Castors
- Tempered Glass Side Wall
- Exhaust Air-Break Canopy Transition
- Arm Support Height Adjustable Foot Rest
- 3 Pieces Work Space
- Motorized Up & Down Door
- IQ/OQ Validation Documentation at Factory
Specification
Model |
LCB-0103B-B2 |
LCB-0123B-B2 |
LCB-0153B-B2 |
LCB-0183B-B2 |
Dimmension Inner (WxDxH mm) Outer |
900x620x680 1050x800x1700 |
1200x620x680 1350x800x1700 |
1500x620x680 1650x800x1700 |
1490x600x635 1950x800x1700 |
Max. door open |
540 mm |
|||
Working door Open |
200 mm |
|||
Air Volume |
34 m³/min |
50 m³/min |
57 m³/min |
68 m³/min |
Filter Type |
Standard: ULPA
U15 Filter, Aluminum framed particle removal 99.999% ISO Class 3, Class 10 HEPA Filter,
& Pre Filter Option – PTFE ULPA
Filter |
|||
Air Flow
Downflow
Inflow |
Standard average
0.3 m/sec, Range 0.25-0.35 m/sec Standard average
0.5 m/sec, Range 0.45-0.65 m/sec |
|||
Noise Sound Level |
Standard <60
dB, Range 56-65 dB(A) |
|||
Illumination |
FL Lamp 20W x
1EA |
FL Lamp 40W x
1EA |
||
UV Disinfection |
UV Lmap 20W x 1
EA |
UV Lmap 40W x 1
EA |
||
Utility Device |
Npt ¼ Air/ Gas
Valve, Vacuum valve, 110V or 220V Electric Socket (Selectable), PAO Valve |
|||
Material
Interior Exterior Door |
Stainless steel
grade 304 with hair lined finish Standard-Epoxy
powder coated electro galvanized steel-Option Stainless steel 304 Slope 7” T.5mm tempered
Safety Glass Sliding Door |
|||
Electric Supply |
110V 60Hz or
220V 50/60Hz |
#Labtech, #Daihanlabtech, #Trias, #pt.trias, #alatlab, #laboratorium, #alatkesehatan, #Daihanlab, #alatindustri, #Laminar Flow, #Fume Hood, #Bench Top Class Ⅱ Type B2, #Bio Hazard Safety Cabinet, #Biohazard Safety Cabinet,