Steam Sterilizer
Feature
Specificaion
MODEL |
LAC-1011SV |
LAC-1021SV |
LAC-1031SV |
Dimensions Inner (WxDxH
mm) Outer |
410 x 560 x 345 200 x 300 x 200 |
490 x 655 x 460 270 x 405 x 230 |
510 x 755 x 545 300 x 455 x 300 |
Capacity |
12 Liter |
25 Liter |
40 Liter |
Heater |
2 kW |
2.5kW |
3kW |
Chamber Type |
Reinforced and Electro polished square chamber |
||
Material |
Stainless Steel 304 standard (STS 316 L option) |
||
Door Lock Mechanism |
Automatic wheel inter locking with one touch sliding
handle |
||
Vacuum Pump |
Built-in diaphragm vacuum pump 25 LPM capacity |
||
Controller |
Microprocessor PID Program Controller with LCD Monitor |
||
Sterilization Temp. Range Pressure Range Course |
110°C - 135°C 0.4-2.3kgf/cm2 Unwrapped 1&2/ Wrapped 1&2/ Liquid/ Bowie-Dick/
User Program mode |
||
Reservoir |
6 L stainless steel tank with top water filling
inlet and silicon closer |
||
Safety |
Over pressure release valve, Over temperature Visible and Audible alarm system, Automatic dual door
locking system |
||
Tray |
180 x 260 mm |
250 x 360 mm |
290 x 400 mm |
Wire tray (Standard) / Perforated tray (Option) |
|||
Electrical Supply |
110V, 60Hz or 220V, 50/60Hz |